How do I correct the Empower 3 error "Unable to store injection" found in the LAC/E32 recover.log file? - WKB47527
Article number: 47527
SYMPTOMS
- Empower 3 error found in the LAC/E32 recover.log file: "Unable to store injection"
- Another error in recover.log: "Failed to recover sample set"
- Buffering indication
ENVIRONMENT
- Empower 3
CAUSE
Low project tablespace; usually less than 10 MB free
FIX or WORKAROUND
- Increase the project tablespace in Configure the System, project properties.
- Close all project-related processes and then exit the project for the change to take affect.