Cannot copy methods in Empower using 'copy to project' function - WKB4116
Article number: 4116
SYMPTOMS
- Cannot copy methods in Empower using 'copy to project' function.
ENVIRONMENT
- Empower 3 personal
CAUSE
Not enough free project tablespace.
FIX or WORKAROUND
- Click on Configure the System to get to Configuration Manager.
- Highlight the destination Project name.
- Right-click and select Properties.
- Increase Quota under "Database tablespace" (minimum of free 10MB required).
ADDITIONAL INFORMATION
id4116, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP