Data missing when reviewing data in Empower - WKB3860

Article number: 3860

SYMPTOMS

  • New system in a different lab than other Empower clients.
  • The location of the new Empower Client is on a new subnet in the company's network.
  • "Data Missing" when trying to review data.

ENVIRONMENT

  • Empower
  • Empower 2
  • Empower 3

CAUSE

The firewall for the router placed between the subnet and that which housed the Empower raw data server was inadvertently turned on.

FIX or WORKAROUND

 Disable the firewall on the router between the two subnets.