No calibration curve data in Empower - WKB3782

Article number: 3782

SYMPTOMS

  • No calibration curve data

ENVIRONMENT

  • Empower 3

CAUSE

Standards are identified as samples.

Standards are identified as control.

FIX or WORKAROUND

  1. Enter Standard for Type, and then reprocess.

ADDITIONAL INFORMATION

 

id3782, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP