Instrument failure, communication failure with ACQUITY Arc - WKB301596

Article number: 301596

SYMPTOMS

  • Instrument failure
  • Communication failure
  • Method setup failure
  • Modules not correctly populated in the WatersDHCP
  • Errors occur starting or during acquisition
  • Full reboot of LAC/E and modules may resolve briefly
  • LAC/E also observed to show black screen be unresponsive

clipboard_e236d96efe2f4e386e42f288f5c912515.png

ENVIRONMENT

  • Empower
  • LAC/E
  • ACQUITY Arc

CAUSE

Corp Network cable is connected to one of the ports of the Instrument LAN Ethernet bridge.

FIX or WORKAROUND

Remove the Corp Ethernet cable from the instrument LAN bridge.