Agilent GC instrument failure with error message in Empower: "The connection to an analysis signal was lost during the run, so the sequence was aborted" - WKB301584
Article number: 301584
SYMPTOMS
- Error message in Empower message center: "InstrumentAgGC#XXXXX:The connection to an analysis signal was lost during the run, so the sequence was aborted", followed by Instrument Failure AgGC#XXXXX
- Error message in Empower message center: "InstrumentAgGC#XXXXX:Could not get control of Agilent XXXX; already controlled by SYSTEM @ XXXXX"
ENVIRONMENT
- Empower 3 Enterprise
- Agilent ICF SL v3.6 and above
- Agilent GC
CAUSE
It is observed on the acquisition client that the Instrument LAN network settings are reset when a Windows time service update is applied.
This leads GC to monitor state and the symptoms noted above.
FIX or WORKAROUND
Stop the time-service updates in a client.