Injection continues for indefinite run time in Agilent 8890 GC, with error "Instrument Failure - End of run wait timed out" - WKB301060

Article number: 301060

SYMPTOMS

  • The sample set contains equilibrate followed by the Inject samples function
  • The control panel shows the run time as 999.9 minutes while running the equilibrate function; does not move to the next injection
  • Issue occurs even if the Agilent 8890 GC firmware is updated above 2.4 

ENVIRONMENT

  • Agilent GC 8890
  • Agilent Instrument Control Framework
  • Empower 3 Enterprise

CAUSE

The ICF support layer is not able to handle the run time for a non-inject function.

FIX or WORKAROUND

This issue is resolved in WICF 4.1 and above (Waters CRI-8097, Agilent WICF-986, and Agilent Internal GC Driver tracking number TT298143).

Note: GC Driver tracking number TT298143 can be seen in Agilent WICF 4.1 release notes.

ADDITIONAL INFORMATION

For Agilent WICF 4.1 driver-related issues, contact the Agilent team directly.

ICF - Instrument Control Framework; ICF SL - Waters ICF Support Layer; WICF - Agilent ICF Support Layer for Waters CDS