In Empower, calibration curve points are missing from the calibration curve - WKB29810
Article number: 29810
SYMPTOMS
- Calibration points are missing from the calibration curve
ENVIRONMENT
- Empower 3
- Calibration curve
CAUSE
The user altered the sample set via the alter sample window in the project browser, instead of from the running sample window, causing the standard injections to become dissociated from the sample set.
FIX or WORKAROUND
Create a process-only sample set.
ADDITIONAL INFORMATION
Ensure that a running sample set is altered only via the running sample window, not from the alter sample window in the project. A warning message appears when a user attempts to alter a sample set via the project.