Skip to main content
Waters

ACQUITY QDa communication failure in Empower - WKB29773

Article number: 29773

SYMPTOMS

  • QDa does not appear in the ACQUITY Console window
  • Cannot reset the QDa from the Empower Run Samples window
  • No data for the QDa in the Run Samples window

ENVIRONMENT

  • Empower 3
  • ACQUITY QDa Mass Detector
  • Other ethernet instruments

CAUSE

Lost communication

FIX or WORKAROUND

  1. Open "Configure the System" from the main Empower window.
  2. Highlight "Nodes".
  3. Highlight the node that the QDa is connected to.
  4. Right-click on the highlighted node and choose "properties".
  5. Click on the "Configure DHCP" tab.
  6. Click the "Configure DHCP" button.
  7. Highlight the entry in the table for the QDa.
  8. Click the "Remove" button below the table.
  9. Power-off the QDa.
  10. Close all Empower windows and log out of Empower.
  11. Reboot the node connected to the QDa.
  12. Once the node reboots, power-on the QDa and wait 6 minutes.
  13. Log in to Empower and repeat steps 1 through 6.
  14. Verify that the QDa is again listed in the table.
  15. Close all Empower windows except the main window.
  16. Open Run Samples and start controlling the QDa.

ADDITIONAL INFORMATION

This is only one solution for a communication failure. There may be others.

Not able to find a solution? Click here to request help.