Component summary table in an Empower report displays information for retention time instead of migration time - WKB29048
Article number: 29048
SYMPTOMS
- Component summary table shows information for retention time instead of migration time for Capillary Electrophoresis (CE)
- Retention time instead of migration time is listed in the title of the table
- Table properties list "original name" as retention time
- CE option is enabled in the project properties
ENVIRONMENT
- Empower
- Capillary Electrophoresis (CE)
CAUSE
Migration time was not selected as the field to summarize for the component summary table.
FIX or WORKAROUND
- Open the report, right-click on the component summary table, and select "table properties".
- Select "component summary tab".
- Under "field to summarize", select migration time. This will change the type of summary for the table. Confirm that the original name changed from "retention time" to "migration time".
- Select Apply > OK.