Inject Samples runs for indefinite run time with Agilent GC; Instrument Failure error in Empower - WKB286729
Article number: 286729
SYMPTOMS
- Inject Samples runs for indefinite run time with Agilent GC; an Instrument Failure error shows in Empower
- End of run wait timed out; no reply from the GC#xxxxx.
- The sample set contains equilibrate followed by the Inject samples function
- The control panel shows the run time as 999.9 minutes while running the equilibrate function
- The first injection after the equilibrate function shows the run time as 999.9 minutes and does not move to the next injection
ENVIRONMENT
- Agilent 7890B
- Agilent 8890
- Agilent Instrument Control Framework
- Empower 3 Enterprise
CAUSE
Firmware issue
FIX or WORKAROUND
- Update the 7890B firmware to 7890B.02.06.1. (Download it from the Agilent website.)
- Update the 8890 firmware to 2.8.1.6. (Download it from the Agilent website.)
- Or, do not use the equilibrate function if the firmware has not been updated.
ADDITIONAL INFORMATION
CRI-3649 - Fixed by Agilent via Firmware update