Article number: 274697

SYMPTOMS

  • Communication error
  • No information about issue in logs
  • Run stops at random time

ENVIRONMENT

  • Empower 3

CAUSE

"Green Ethernet" settings are on for both instrument and domain network cards.

FIX or WORKAROUND

On both domain and instrument network cards, "Green Ethernet" settings should be turned off. See all settings below:1.png

1. Allow the computer to turn off this device to save power should be unchecked (disabled)

2.png

2. Energy-Efficient Ethernet

3.png

3. Green Ethernet

4.png

4. Idle Power Saving

5.png

5. Large Send Offload v2 (IPv4)

6.png

6. Large Send Offload v2 (IPv6)

7.png

7. TCP Checksum Offload (IPv4)

8.png

8. TCP Checksum Offload (IPv6)

9.png

9. UDP Checksum Offload (IPv4)

10.png

10. UDP Checksum Offload (IPv6)

11.png

11. WOL & Shutdown Link Speed

12.png

12. Ultra Low Power Mode

clipboard_bf864719-50cf-4ae0-87d6-dcf5e1fec4b6.png

ADDITIONAL INFORMATION