Skip to main content
Waters

Re-using older calibration curve in Empower - WKB274031

Article number: 274031

OBJECTIVE or GOAL

How to use an older calibration curve to quantitate samples. Every time Standards are calibrated, Empower assigns a Calibration ID to the calibration curves
generated. If one wants to use a calibration curve other than the current one, one need to select the older curve using the Calibration ID number.

ENVIRONMENT

  • Empower

PROCEDURE

1.       In this example, the current calibration has ID 6240 and the older calibration has ID 6213. First look at the difference between them using the Compare window.

clipboard_e530ccc436dc753226b1e90745a61fd7f.png

2.       In Compare Window with Calibration ID 6240 highlighted, click the Individual Points tab

clipboard_efba5a8ccd367a1553473acdfd386b493.png

3.       The X Values for the three levels are 0.02, 0.05, and 0.1 respectively

clipboard_ebd488d4167435fd09714eff10ddf5933.png

4.       The older Calibration ID of 6213 has X Values that are slightly lower due to a Purity factor that was applied. The goal is to use this older calibration curve where the Purity factor had been applied.

clipboard_e600db3c15984106c5429e168d1cb1302.png

5.       Bring a Result Set or a Result into Review and click the Calibration Curve tool

clipboard_efa55d88f2ab3f89d20f91a657e5999cc.png

 

6.       In the Calibration Curve window go to Options, Select Calibration

clipboard_e21d284c5e4e37759e84e67d00262a385.png

7.       Select the Calibration Curve you want to use (in this case Calibration ID 6213) and click OK

clipboard_ea68fde8f2200cb19b713dbd5701398a9.png

8.       Click OK to the warning message that says the calibration you have selected is not the most current

clipboard_ee110b2f9ca8ffa3e9d289783deb494cc.png

9.       Calibration curves are associated with the processing method, so from the File menu select Save As, Processing Method to give the method a different name

clipboard_ebe0d55f1a92f9ed977175580572fd971.png

10.   You will be asked to Clear Curves or Copy Curves. Select Copy Curves

clipboard_e02b4e6902a94e6dbf41e36ea0d0ebfc7.png

11.   Enter a new name for the Processing Method and click Save

clipboard_ef15fb0dd88cf03b7d2b521cb8a52089c.png

12.   Exit from Review and now process the Sample Set using the older calibration curve. Go to the Sample Sets tab in Browse Project, right-click on the Sample Set and select Process

clipboard_e5d0ae60001cabd1842e6e1fa8fed0254.png

13.   In the Background Processing and Reporting dialogue box select the Processing Method with the different name, and from the How drop-down menu select Quantitate Only. Then click OK.

clipboard_eae8d3e9cbc3707d74a653523a7dac795.png

14.   Go to the Result Sets tab, right-click on the new Result Set and select Review.

clipboard_e018a72388959b40e4e929b6c498723df.png

 

15.   Click the Calibration Curve tool.

clipboard_e71d83ada3d7cfc293222d6876713c281.png

16.   As seen in our example, the X Values for Uracil are from the older calibration curve.

clipboard_e98d3fa5c540573586e4b1cbf2e2e2954.png

17.   If we compare the results for a sample, we see a different calculated amount for Uracil when we use the older calibration curve.

clipboard_ed3535a40da937f60ebffc09b3cec2cf3.png

ADDITIONAL INFORMATION

 

id274031, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.