Skip to main content
Waters

Intermittent communication failure in ACQUITY console - WKB273444

Article number: 273444

SYMPTOMS

  • Intermittent communication failure in ACQUITY console

ENVIRONMENT

  • Empower

CAUSE

Both the Instrument LAN and Corporate LAN IP addresses start with 192.168...

FIX or WORKAROUND

Use the alternative IP address (172.16.0.1) for the LC-MS system.

ADDITIONAL INFORMATION

 

id273444, comms

Not able to find a solution? Click here to request help.