Intermittent communication failure in ACQUITY console - WKB273444

Article number: 273444

SYMPTOMS

  • Intermittent communication failure in ACQUITY console

ENVIRONMENT

  • Empower

CAUSE

Both the Instrument LAN and Corporate LAN IP addresses start with 192.168...

FIX or WORKAROUND

Use the alternative IP address (172.16.0.1) for the LC-MS system.

ADDITIONAL INFORMATION