Instrument Failure - End of run wait timed out at xxxx seconds. No reply from W2489 - WKB272254

Article number: 272254

SYMPTOMS

  • Instrument Failure - End of run wait timed out at 4201.1 seconds. No reply from W2489#serialnumber
  • SQT performed without issue on 12 new Arc systems
  • Three were put into use and all showed the issue
  • Issue seen on all sample sets after one or two injections

ENVIRONMENT

  • Empower
  • FR5
  • DP2018R1
  • Customer-supplied HP Elite SFF 600 G9

CAUSE

Customer found:

  • NIC drivers were 2015.
  • Multi-Core support was disabled.

FIX or WORKAROUND

  1. Update the NIC drivers.
  2. Enable Multi-core support.

ADDITIONAL INFORMATION

Note: Waters-supplied Config 16 LAC/E Multi-Core support is a BIOS option - Enable Multi-Core Support (it is enabled by default)