Instrument Failure - End of run wait timed out at xxxx seconds. No reply from W2489 - WKB272254
Article number: 272254
SYMPTOMS
- Instrument Failure - End of run wait timed out at 4201.1 seconds. No reply from W2489#serialnumber
- SQT performed without issue on 12 new Arc systems
- Three were put into use and all showed the issue
- Issue seen on all sample sets after one or two injections
ENVIRONMENT
- Empower
- FR5
- DP2018R1
- Customer-supplied HP Elite SFF 600 G9
CAUSE
Customer found:
- NIC drivers were 2015.
- Multi-Core support was disabled.
FIX or WORKAROUND
- Update the NIC drivers.
- Enable Multi-core support.
ADDITIONAL INFORMATION
Note: Waters-supplied Config 16 LAC/E Multi-Core support is a BIOS option - Enable Multi-Core Support (it is enabled by default)