Clear Calibration function does not work as expected in Empower - WKB264325
Article number: 264325
SYMPTOMS
- When batch processing a Sample Set with several Clear Calibration lines and expecting a new calibration curve to be created after each Clear Calibration function, only a single calibration curve is created with each additional standard included as new points on the curve
ENVIRONMENT
- Empower 3
- 2489/2487 UV/Vis Detector
CAUSE
Data channel is a 2487, but the instrument method in the Method Set has a 2489 instead of a 2487.
FIX or WORKAROUND
- Remove the 2489 from the Instrument Method and add a 2487.
- Reprocess the Sample Set.
ADDITIONAL INFORMATION
- "Clear Calibration" line will not execute if the sample set starts in "Run Only" mode.
- Sample set contains first row "Clear Calibration" and Processing type as "Normal".
- Sample set also contains "Clear Calibration" at the end of all the injections, but the processing type selected is "Don't process or report", so this line will be ignored.