Processing issues from RDPS Microsoft Virtual Desktop (VDI) clients - WKB263931

Article number: 263931

SYMPTOMS

  • Processing issues

ENVIRONMENT

  • Empower
  • Microsoft Virtual Desktop Infrastructure (VDI)
  • RDPS clients
  • Windows Server 2019

CAUSE

The WatersService and Waters DHCP service are not disabled on the VDI servers.

FIX or WORKAROUND

Disable the WatersService and Waters DHCP service on VDI servers (as is required for Citrix).

ADDITIONAL INFORMATION

Microsoft Virtual Desktop Infrastructure (VDI) is not officially tested with Empower.