Error: Invalid curve with GPC data and Empower - WKB26236

Article number: 26236

SYMPTOMS

  • "Invalid curve" message in the calibration window
  • Sample type is set to Broad Unknown and Narrow Standards
  • Vo and Vt values are set up correctly

ENVIRONMENT

  • Empower 3

CAUSE

Standards are running using a different instrument method than the unknowns.

FIX or WORKAROUND

  1. Rerun the standards and unknowns using the same instrument method.
  2. Process and save calibration .
  3. Process the unknowns.

ADDITIONAL INFORMATION