From Citrix server, cannot control any Agilent ICF module from the Empower Run Samples window - WKB254661
Article number: 254661
SYMPTOMS
- From Citrix server, cannot control any Agilent ICF module from the Empower Run Samples
ENVIRONMENT
- Empower
- Agilent ICFSL v3.6.0/3.6.1
- Citrix
CAUSE
ICF Support Layer 3.6.1 and 3.6.0 are not backward or forward compatible. ICF Support Layer 3.7.01 Release notes have details on the compatibility.
FIX or WORKAROUND
Use the same Agilent Support Layer version on Citrix and LAC/Es.
This issue is not observed when ICF Support Layer 3.5.1 or below is installed on a LAC/E device and ICF Support Layer 3.7.01 is installed on a client / Citrix Server.
ADDITIONAL INFORMATION