Why are the molecular weights missing for the standard injections in the Empower 3 GPC Review window? - WKB25264
Article number: 25264
SYMPTOMS
- After processing GPC data, there are no molecular weight values for the standard injections
ENVIRONMENT
- Empower 3
- GPC analysis
CAUSE
The molecular weights were not entered manually by the user in the component editor window.
FIX or WORKAROUND
- Highlight the sample set that was run.
- Right-click it.
- Select Alter Sample.
- In the Alter Sample window, click Edit in the toolbar and select Amount. This opens the Component Editor.
- In the table, specify the known molecular weights for the standards.
- When all of the molecular weights are specified, click OK.
- In the Alter Sample window, click File > Save.
- Close Alter Sample and reprocess the sample set.
ADDITIONAL INFORMATION
In the Alter Sample window, be certain that the Sample Type for the standards is set to Narrow Standard.