Error message "Attempted an unsupported operation" when trying to edit a Dissolution Table report group - WKB25185

Article number: 25185

SYMPTOMS

  • When attempting to edit a Dissolution Table report group, an error message appears: "Attempted an unsupported operation"
  • The report group can be edited, but changes cannot be saved for Bath Grouping or Use Exponential Notation properties.
  • Although the error message may appear twice when making selections in the dialog, changes to other properties save successfully.

ENVIRONMENT

  • Empower 3 FR3
  • Empower 3 FR3 Hotfix 1

CAUSE

A defect was introduced with the release of Empower 3 FR3: PCS# 54838 - Dissolution Table Report Group - Can't Set Bath Grouping.

FIX or WORKAROUND

PCS# 54838 was fixed in Empower 3 FR4.

ADDITIONAL INFORMATION

  • Using Empower 3 SR2, two report methods were saved that can be dropped into an Empower 3 FR3 project to get a Dissolution Table report group with baths A and B in separate tables or in the same table.
  • This gives users the option of setting up a Dissolution Table with each type of bath grouping.
  • To get a copy of each Report Method file, go to the Attachments section below, right-click and select "Save target as ..." to save the MRP file.
  • These files cannot be opened, but you can use drag-and-drop to put them into an Empower 3 FR3 project.