Empower run does not start if Wet Prime function is present in the sample set method - WKB248590
Article number: 248590
SYMPTOMS
- Empower run does not start if Wet Prime function is present in sample set method
- A value is required in the "Sample Prep" column
ENVIRONMENT
- Empower 3 version 3.7.0
- Alliance e2695
CAUSE
Unknown
FIX or WORKAROUND
Specify the value 1 in the "Sample Prep" column for the row that includes the Wet Prime function.