Processing through channels generates no results for PDA data - WKB247163

Article number: 247163

SYMPTOMS

  • No results generated when processing through channels
  • Results are partly generated when processing through injections, but without any integration
  • Acquisition method set is used for processing 

ENVIRONMENT

  • Empower 3

CAUSE

Extracted 2D channels from a 3D chromatograms were not saved.

FIX or WORKAROUND

Use the "Save PDA Extracted Channels" option in the method set.