Empower cannot process Perkin Elmer data with Clear Calibration function - WKB239627

Article number: 239627

SYMPTOMS

  • Cannot process data collected from a PE system (this applies, apparently, to various instrumentation)
  • Error message: "No Channels to clear from Method Set"
  • Method set has an instrument method specified
  • Channel names are correctly reported in the method set/processing method

ENVIRONMENT

  • Empower 2
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise
  • Perkin Elmer instrument

CAUSE

Channel is not populated automatically in method set when PE driver is used. 

FIX or WORKAROUND

Remove the Clear Calibration line from the sample set itself, and use the Clear Calibration option when performing batch processing. If the sample set was already acquired, perform Alter Sample from the Sample Set tab in the Project window.

ADDITIONAL INFORMATION

See also wkb94191 for the same behavior with Thermo data.

 