Dates are displayed incorrectly in Empower 3 - WKB238190

Article number: 238190

SYMPTOMS

  • Empower 3 shows incorrect date formats

clipboard_e996b4c04e425a64ee7f2d7402e15e821.png

ENVIRONMENT

  • Empower 3 

CAUSE

In the Windows Region, Additional settings, the Date formats for "Short date" and "Long date" are incorrectly specified.

clipboard_e15abbaea1e5c9293f3436ffbf6e0aad0.png

FIX or WORKAROUND

  1. Specify the correct Date formats for "Short date" and "Long date".

ADDITIONAL INFORMATION

The Windows operating system date format will apply to Empower software. The Empower server and all Empower clients and LAC/Es must be on the same date format.