With Agilent ICF Support 3.4, HS 7697, and GC 8860, second injection stays in preparing for injection mode, and then a Server Busy error message appears - WKB235874
Article number: 235874
SYMPTOMS
- HS 7697, GC 8860, front injector selected in the instrument method (no issue in headspace mode)
- First injection is fine, but second injection stays in Preparing for injection mode, and then Server Busy Retry/Switch to message appears
- Must restart the LAC/E and the GC to be able to control the GC again
ENVIRONMENT
- Empower 3
- Agilent ICF Support 3.4
- Agilent 8890
- Headspace
CAUSE
J&W is selected as Manufacturer in additional information for the GC column.
FIX or WORKAROUND
In column configuration > Manufacturer, specify Agilent instead of J&W.