Why does Empower display "Data Incomplete" in the Review window? - WKB23229

Article number: 23229

ENVIRONMENT

  • Empower 3

ANSWER

When Empower starts an injection, it looks at the runtime and the data acquisition rate for the detector being used. This tells it how many data points it should have in total for that data channel, once it is collected. When the run is finished and the channel is taken into Review, Empower shows "Data Incomplete" if the total number of data points in the chromatogram does not equal the number that it calculated before starting the run.