Agilent 7890B error "Information: Method: Configuration has changed. The following parts of the method are not consistent with the GC configuration . . ." - WKB230859

Article number: 230859

SYMPTOMS

  •  Agilent 7890B error - Information: Method: Configuration has changed. The following parts of the method are not consistent with the GC configuration: Item method GC Column 1 Temperature 60 degrees C - 240 degrees C (240 degrees C) 0 degrees C - 240 degrees C (240 degrees C)

ENVIRONMENT

  • Agilent 7890B GC
  • Empower 3.6.1
  • ICF SL 3.4

CAUSE

Incorrect information in the instrument method and lack of communications with the GC

FIX or WORKAROUND

Power cycle the GC and LAC/E32, and then, in the GC instrument method, install the catalog information for the column manufacturer and column part number. Ensure that the minimum temperature is set accordingly.

ADDITIONAL INFORMATION

 