Empower 3 - Issue in Review > open processing method > Error > fails to start application application (Processing Method Editor) - WKB228627
Article number: 228627
SYMPTOMS
- Issue In Review > Open processing method - Window closes > Method does not open
- Fails to start application (Processing Method Editor)
ENVIRONMENT
- Empower 3
CAUSE
Project has a custom field that references itself.

FIX or WORKAROUND
Remove the circular reference.
ADDITIONAL INFORMATION