Is it possible to set privileges in Empower 3 that allow automatic processing but prevent manual processing? - WKB225438
Article number: 225438
ENVIRONMENT
- Empower 3
ANSWER
1. Automatic processing results can be saved, but manual analysis results cannot be saved.
- Turn on the privilege "Save Results" and turn off the privilege "Save Results & Calibrations in Review".
2. Both automatic processing results and manual processing results can be saved.
- Turn on both the privilege "Save Results" and the privilege "Save Results & Calibrations in Review".
"Automatic processing results" refers to the results that are processed by right-clicking on a sample set in the project window.
"Manual processing results" refers to the results that are processed and saved in the Review window.
ADDITIONAL INFORMATION