Some injections are processed and associated to a result set, but others are not - WKB225204
Article number: 225204
SYMPTOMS
- Only some injections are processed
- Each injection is processed immediately after acquisition completes
- No audit trail for the result set creation
ENVIRONMENT
- Empower
- Run Only
- Run & Process
- Run & Report
CAUSE
Users can change the run mode during acquisition.
The run mode and changes to the run mode are not audit trailed.
FIX or WORKAROUND
Do not change the run mode during acquisition.
ADDITIONAL INFORMATION
- There is an enhancement request to implement changes to prevent this: https://code.waters.com/jira/browse/CRI-1945
- There is an enhancement to record the run and suitability modes: https://code.waters.com/jira/browse/CRI-561
- See WKB224214 and WKB118203