Processing stops responding on Empower running with Windows Virtual Desktop (WVD) application servers - WKB219482

Article number: 219482

SYMPTOMS

  • Processing intermittently stops responding 
  • Processing stops 
  • Later attempts to process are successful
  • Clearing the processing queue may resolve the issue, but it often does not help

ENVIRONMENT

  • Empower
  • Windows Virtual Desktop (WVD) application servers
  • Azure Cloud deployment

CAUSE

The WatersService was not disabled on the Windows Virtual Desktop servers, causing queuing of the processing jobs.

FIX or WORKAROUND

As directed in the Empower installation guide (ICUG) for Citrix deployments: Disable Waters Service and Waters DHCP Server on the Citrix application servers.

ADDITIONAL INFORMATION

See page 144 of the Empower 3.6.1 ICUG.