Empower Agilent G1365A error: Data invalid. At least for some diodes of the array the data produced are invalid due to too much light at a diode or a gain setting - WKB21646
Article number: 21646
SYMPTOMS
- Agilent G1365A produces the error message "Data invalid. At least for some diodes of the array the data produced are invalid due to too much light at a diode or a gain setting."
ENVIRONMENT
- Empower 3
- Agilent
CAUSE
The margin for negative absorbance is set too low.
FIX or WORKAROUND
- Change the parameter "Margin for negative absorbance" to a higher value.
ADDITIONAL INFORMATION
It is possible that negative absorbance is high at the beginning of the run, or it may be just a reference issue.
The reference spectra may need to be gathered from another part (a quieter part) of the chromatogram.
One case required an increase in value to 500 mAU.