"System Idle - Instrument Failure" error message for Agilent 8890 GC with liquid sampler in Empower - WKB207718

Article number: 207718

SYMPTOMS

  • "System Idle - Instrument Failure" error message for Agilent 8890 GC with liquid injection appears after runtime plus 10 minutes
  • The sample set has Equilibrate function, followed by Inject Standards/Samples

ENVIRONMENT

  • Empower 3 Enterprise
  • Agilent Instrument Control Framework (ICF SL v3.5)
  • Agilent 8890 with ALS

CAUSE

The column manufacture field is empty on the Columns tab.

clipboard_ebd5fb2d643e50a23c8dd59644d462503.png

FIX or WORKAROUND

From the drop-down list, select the column manufacture as Agilent or J&W, and save the instrument method.