Skip to main content
Waters

How to exclude points from a calibration curve and reprocess data in Empower - WKB20751

Article number: 20751

OBJECTIVE or GOAL

Exclude points from a calibration curve and reprocess data in Empower.

ENVIRONMENT

  • Empower 3

PROCEDURE

  1. Go to the Results tab and bring results into Review.
  2. Open the Calibration curve window.
  3. Ignore point(s) (Refer to How do I ignore a point in an Empower calibration curve?) 
  4. Process > Calibrate.
  5. File > Save Calibration, and then close Review.
  6. Open the Project window.
  7. Select Result Set > Process > Quantitate Only.
  8. Select Result Set, double-click to Results, see the latest results or select Result Set and bring to Review, and view Results to confirm proper 
    calculations.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.