Empwer error message "Method download failed because module is offline!" for Agilent LC with ICFSL v3.5 - WKB207291

Article number: 207291

SYMPTOMS

  • The error message "Method download failed because module is offline!" appears in the Empower message center when running an Agilent LC
  • Agilent LC instrument module temporarily goes into offline state

ENVIRONMENT

  • Agilent ICFSL v3.5

CAUSE

With Agilent LC Driver 3.3 (part of ICFSL v3.5), due to error handling, new locks are introduced at the driver level. This results in frequent occurrences of this problem.

FIX or WORKAROUND

Fixed in ICF SL 3.5.1

ADDITIONAL INFORMATION

CRI-3716 was raised for this issue.

The Agilent KPR number for this defect is 574916.

Same issue reported with Agilent ICFSL v3.6, CRI-7322 was raised for this issue.