Cannot calculate USP tailing with Empower 3 - WKB20522
Article number: 20522
SYMPTOMS
- USP tailing is missing for a peak with shoulder on right-hand side
- Peak code is S06
ENVIRONMENT
- Empower 3
CAUSE
S06 code means there is a problem calculating peak width at 5% height (which is needed for USP tailing factor).
FIX or WORKAROUND
1. Try exponential and tangential non-classic front skim.
2. If that does not work, optimize the method for better separation.
ADDITIONAL INFORMATION
id20522, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP