In Empower, a bad checksum message appears after verifying incomplete data - WKB204954
Article number: 204954
SYMPTOMS
- Bad checksum message appears after verifying incomplete data
ENVIRONMENT
- Empower 3 FR1 to FR 5, SR5
CAUSE
The checksum was not properly updated when a saved derived channel was closed.
Opened as CRI-2391/CRI-703
FIX or WORKAROUND
Waters has improved the ability to diagnose this problem in Empower 3 FR4 SR3. In this scenario, the Channel Status would be "Being Acquired" and the Injection Status would be "Incomplete". The user can verify the data as complete.
ADDITIONAL INFORMATION
Assigned to Engineering as Priority 1.
When resolved, it will be published in Release Notes.