Empower does not move to next injection - WKB204259

Article number: 204259

SYMPTOMS

  • System runs first injection but does not move to the next injection

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2998 Photodiode Array Detector
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise

CAUSE

TCP/IPv6 is turned on.

FIX or WORKAROUND

  1. Clear the TCP/IPv6 selection from network adaptor properties.

clipboard_efcb0f922ec4a27ca5f1b5f881f659872.png

  1. Also, disable TCP/IPv6 from Registry. See WKB9503 for instructions on disabling TCP/IPv6.

ADDITIONAL INFORMATION

 