Empower does not move to next injection - WKB204259

Article number: 204259

SYMPTOMS

  • System runs first injection but does not move to the next injection

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2998 Photodiode Array Detector
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise

CAUSE

TCP/IPv6 is turned on.

FIX or WORKAROUND

  1. Clear the TCP/IPv6 selection from network adaptor properties.
  2. Also, disable TCP/IPv6 from Registry. See WKB9503 for instructions on disabling TCP/IPv6.

ADDITIONAL INFORMATION

 

id204259, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, 2998, 2998B, 2998SFC, A-10PDA, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP