In Empower, short run time observed using 2695 and 2996 - WKB20302
Article number: 20302
SYMPTOMS
- Short run time observed in sample set (for example, run time is set to 30 minutes but sample runs only to 27 minutes)
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2996 PDA Detector
- Empower 3
CAUSE
This is a known defect (PCS 53018) where there is potential for an instrument to reset its own sampling rate, resulting in shorter run times.
FIX or WORKAROUND
None. This is a rare occurrence.
Waters recommends that chromatograms be visually inspected for anomalies, such as an actual run time differing from the expected run time.
ADDITIONAL INFORMATION