In Empower, short run time observed using 2695 and 2996 - WKB20302

Article number: 20302

SYMPTOMS

  • Short run time observed in sample set (for example, run time is set to 30 minutes but sample runs only to 27 minutes)

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2996 PDA Detector
  • Empower 3

CAUSE

This is a known defect (PCS 53018) where there is potential for an instrument to reset its own sampling rate, resulting in shorter run times.

FIX or WORKAROUND

None. This is a rare occurrence.

Waters recommends that chromatograms be visually inspected for anomalies, such as an actual run time differing from the expected run time.

ADDITIONAL INFORMATION

 

id20302,