Article number: 201288

ENVIRONMENT

  • Empower 3

ANSWER

- Normal: Follow the contents of the Run mode specified at the start of the sample set.

- Don't Process or Report: Run in "Run and Process" or “Run and Report” mode. The sample set is acquired, but no processing or reporting is preformed for this row

- Don’t Report: Do not report this row (run the processing), even if the sample set is run in "Run and Report" mode.

- Ignore Faults: Continue injections even if a fault condition is detected. When "Stop on Fault" or "Reinject on Fault" is specified in the Suitability mode at the start of the sample set, it is ignored for this row.

- Ignore Stability: Disables the settings on the "Suitability" tab of the processing method. The Suitability mode settings are enabled. For example, if "Stop on Fault" is specified in Suitability mode and "Must peak" is registered in the processing method, the acquisition stops when the "Must peak" is not detected.

 