Display of the instrument method AgilentLC is truncated when using Empower with Microsoft Remote Desktop connection - WKB193874

Article number: 193874

SYMPTOMS

  • The display of the instrument method AgilentLC is truncated when using Microsoft Remote Desktop connection

    clipboard_eecbceee024c15a2aeedf686998ffd739.png
  • No issue from a LAC/E32

ENVIRONMENT

  • Empower Enterprise
  • Agilent ICF Support 3.4
  • Microsoft Remote Desktop Connection

CAUSE

Screen size/resolution
Waters does not test and therefore does not not officially support running Empower via Microsoft Remote Desktop Connection.

FIX or WORKAROUND

  1. Try to change in Scale and Layout "Change the size of text, apps, and other items" in Windows settings.

    Select 100% or try other values.
    clipboard_ed777c651e989067c91b1e38fefd12c4d.png
  2. Use a bigger screen with another resolution.
  3. Use a 4:3 display resolution

ADDITIONAL INFORMATION

 