8890/7697A instrument failure after ICF Support Layer 3.4 upgrade - WKB193233
Article number: 193233
SYMPTOMS
- Instrument failure in message center
- Agilent Pre-Configuration Tool successfully populates instrument configuration
- IP address configured correctly as static on the front panel of the GC and Headspace
- IP addresses for both GC and Headspace communicate successfully
- Instrument information configuration correct in DHCP
ENVIRONMENT
- Agilent 8890 GC
- Agilent 7697A Headspace Analyzer
- Empower 3
- ICF Support Layer 3.4
CAUSE
Agilent drivers A.02.04 on LAC/E
FIX or WORKAROUND
Update drivers on the LAC/E to match those in ICF SL 3.4.