8890/7697A instrument failure after ICF Support Layer 3.4 upgrade - WKB193233

Article number: 193233

SYMPTOMS

  • Instrument failure in message center
  • Agilent Pre-Configuration Tool successfully populates instrument configuration
  • IP address configured correctly as static on the front panel of the GC and Headspace
  • IP addresses for both GC and Headspace communicate successfully
  • Instrument information configuration correct in DHCP

ENVIRONMENT

  • Agilent 8890 GC
  • Agilent 7697A Headspace Analyzer
  • Empower 3
  • ICF Support Layer 3.4

CAUSE

Agilent drivers A.02.04 on LAC/E

FIX or WORKAROUND

Update drivers on the LAC/E to match those in ICF SL 3.4.

ADDITIONAL INFORMATION

 

id193233, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP