Agilent LC freezes up while equilibrating under Empower control - WKB1926
Article number: 1926
SYMPTOMS
- When equilibrating in Empower, the Agilent LC system freezes and never progresses
ENVIRONMENT
- Empower Enterprise
- Agilent LC 2.1 Hotfix 1 ICF 1.04 ICF-LC Driver 1.04
CAUSE
Defect in driver
FIX or WORKAROUND
- Upgrade to the current version Agilent LC Driver 2.2.
- Completely remove the Agilent 2.1 through Windows Start Menu > All Programs > Empower Folder > Remove Instrument Component Software logged into the computer as a Local Administrator.
- Remove the ICF and ICF-LC Drivers through Programs and Features (Add/Remove Programs).
- Restart the computer.
- Install Agilent Driver 2.2 as a local Administrator.
- Restart the computer.
- Run as normal.