Empower 3 GC driver 2.6 error when running a 7890B GC: There is an error in XML document (1, 1900) - WKB19099
Article number: 19099
SYMPTOMS
- When Run Samples is opened, the following message appears: "There is an error in XML document (1, 1900)" (number may be different) and the instrument does not run.
- Message center full with Error Instrument: "There is an error in XML document (1, 1900)"

-
There is an error in XML document (1,1892)
-
There is an error in XML document (1,1888)
-
There is an error in XML document (1,1890)
ENVIRONMENT
- 7890B
- Empower
- Instrument Control Software 2.6
- Firmware B.02.05 , B.02.04.3, B.02.03.1, B.02.05
CAUSE
In the control panel of the GC 7890B, Sleep Mode was selected for back inlet (back inlet was not selected in the instrument method).
FIX or WORKAROUND
- Change the mode for the back inlet from Sleep to something else (Split or Splitless, for example).
- Do not use Sleep mode for the inlet on a GC 7890B controlled by Empower with GC driver 2.6.
ADDITIONAL INFORMATION
In this case, downgrade firmware did not solve the problem.
7890 firmware B.02.05 seems to be ok.
