Special default custom fields cannot be copied to project - WKB190727
Article number: 190727
SYMPTOMS
- Cannot copy special default custom fields to a project in the project window
ENVIRONMENT
- Empower
- Default Custom Fields
- ColumnName
- ColumnSerial Number
CAUSE
Special or reserved custom fields cannot be copied to projects. This includes certain SQT custom fields and the two default fields containing Column.
FIX or WORKAROUND
- From the custom field tab of the project properties use the 'Move to Project' function.
- Use the clone project function
ADDITIONAL INFORMATION
Also see WKB80687 regarding the special, hidden SQT custom fields.
NOTE: The default custom fields are in every project and cannot be deleted therefore when saving them to a project there is no audit trail recorded.
id190727, EMP2LIC, EMP2OPT, EMP2QLIC, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3Q1ULIC, EMP3QLIC, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP