Waters DHCP Configuration screen is missing Type and Name - WKB16322
Article number: 16322
SYMPTOMS
- Unable to connect to instruments
- Ipconfig/all shows the Instrument LAN IPv4 address correctly
- Pinging the Instrument LAN adapter returns the correct IPv4 address with no IPv6 addresses in the list
- IP address and MAC address are listed for all instruments
- Only Name and Type are blank
ENVIRONMENT
- Empower 3 Base and higher
- AVAST Business Antivirus package installed
- Other antivirus packages may also have this feature (360 Total Security)
CAUSE
The antivirus package contains a firewall that is blocking the return of Name and Type.
FIX or WORKAROUND
- Open the AVAST software from the task bar on the PC that has the instruments.
- Select Firewall.
- The Instrument LAN will show as an unidentified network with public access.
- Change the access from Public to Private by sliding the bar to Private.
- Name and Type should now appear.