Agilent 6890 Empower 3 SR2 error: Can only run dual tower enabled method in dual tower mode - WKB1430
Article number: 1430
SYMPTOMS
- When attempting to load a method on a GC, the user encounters the error message "Can only run dual tower enabled method in dual tower mode".
ENVIRONMENT
- Empower 3 SR2 and later
- Agilent 6890 GC with single tower ALS
CAUSE
- Beginning with Empower 2 FR5, the F:/B: prefix to vial numbers was required because of added supported for dual tower mode without dual towers installed.
- With the release of Empower 3 SR2, there was another change. The F:/B: prefix are only required if two towers are installed. Otherwise, you only need vial number.
FIX or WORKAROUND
- Add a second tower or alter the sample sets to use the correct vial syntax.
- It is necessary to create new methods. Existing methods cannot be modified.
ADDITIONAL INFORMATION
See Designating vial numbers for 6890 GC samples (PCS54498)
id1430, 6890A/+GC, 6890N GC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP