Skip to main content
Waters

Data processing in Empower creates multiple new calibration curves instead of one curve per component - WKB12716

Article number: 12716

SYMPTOMS

  • Data processing creates 20 new calibration curves instead of the typical 4 for four components

ENVIRONMENT

  • Empower

CAUSE

In the processing method, on the Components tab, "Replace" was selected in the Update RT field. The Update RT field adjusts the retention time of calibration curves. Selecting "Replace" updates the retention time of the calibration curve every time a chromatogram is calibrated or quantitated, regardless of the sample type.

clipboard_e690697944407aa54a7aad9a3d1025492.png

FIX or WORKAROUND

Select "Never" for Update RT and reprocess.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.