Can I have a single instrument configuration in Empower and turn components on/off as needed? - WKB124472
Article number: 124472
ENVIRONMENT
- Empower Enterprise
- Empower Workgroup
- Empower Personal
ANSWER
No. Each Empower chromatographic system needs to be created with only the LC or GC modules that will be used for the particular chromatographic analysis.
For multiple detector configurations, you must have separate chromatographic systems for each individual or multiple detector combination, sharing the same sample manager, solvent manager, and column heater/manager. The user then chooses the project and the appropriate chromatographic system as needed for their particular analysis.
ADDITIONAL INFORMATION
For example, if one has a LC system with both a TUV and FLR detector, create three chromatographic systems,
- LC with TUV
- LC with FLR
- LC with TUV plus FLR
and choose the system appropriate for the type of data to be collected.